Product Details
Place of Origin: China
Brand Name: Senova
Certification: CE
Model Number: NuOven B252-I
Payment & Shipping Terms
Minimum Order Quantity: 1 unit
Price: Available upon request
Packaging Details: Carton with pallet
Delivery Time: Approx. 10 working days
Payment Terms: T/T
Supply Ability: 500 units/month
| Model | NuOven B31-I | NuOven B72-I | NuOven B101-I | NuOven B150-I | NuOven B252-I | ||
| NuOven C31-I(*Note) | NuOven C72-I (*Note) | NuOven C101-I (*Note) | NuOven C150-I (*Note) | NuOven C252-I(*Note) | |||
| Chamber volume(L) | 30 | 70 | 100 | 150 | 250 | ||
| Convection Method | Horizontal forced Convection (Fan) | ||||||
| Temp. Control Range | RT+10℃~ 400℃ | ||||||
| Temperature | Accuracy | ±1℃ at 100℃ | |||||
| Fluctuation | ±1.5℃ | ||||||
| Uniformity | ± 1.5℃ at 100℃ | ± 2.5℃ at 100℃ | |||||
| Controller | PID microprocessor control, soft touch, LED display | ||||||
| Sensor | PT100 | ||||||
| Timer | Power-on, power off and working. Timing range: 1min-99hr | ||||||
| Material | Internal | 304 stainless steel | |||||
| External | Steel (powder coating) | ||||||
| Dimensions (WxDxH,mm) | Internal | 300*300*350 | 400*400*450 | 450*450*500 | 500*500*600 | 600*600*700 | |
| External | 640*520*575 | 740*620*675 | 790*670*725 | 840*720*825 | 940*820*925 | ||
| Net Weight(Kg) | 46 | 65 | 78 | 94 | 125 | ||
| Consumption Power(W) | 1200 | 1600 | 2000 | 2500 | 3200 | ||
| Shelf Size(mm) | 290*270 | 390*370 | 440*420 | 490*470 | 590*570 | ||
| ShelfQty(Standard/Maximum) | 2/4 | 2/4 | 2/4 | 2/4 | 2/5 | ||
| Power Supply | 220V/50Hz (Optional: 220V/60Hz, 110V/60Hz) | ||||||
*Note:NuOven C31-I/C72-I/C101-I/C150-I/ C252-I has10-segment temperature control program to realize temperature curve program running.
Many industries operate under strict guidelines (e.g., ISO, ASTM, USP). Laboratory drying ovens used in these settings must comply with specific standards. Protocols will dictate exact drying temperatures and times for given materials. For example, USP <701> for pharmaceutical glassware sterilization specifies time and temperature profiles for dry heat.
Compliance often requires documented evidence of the oven's performance. This includes installation qualification (IQ), operational qualification (OQ) verifying temperature uniformity across the workspace, and performance qualification (PQ) showing it works for the intended process. Regular calibration and preventive maintenance logs are part of this compliance framework, proving the instrument is fit for its regulated purpose.