Product:
Laboratory Water Purification System With EDI
Water Requirement:
Tap Water(TDS<<2000us/cm)
Output Capacity:
30L/hour (Class III),15 L/hour (Class II), 1.5~2.0L/min (Class I)
Dispense Volume:
100-60000ml
Water Tank Capacity:
30L
Control System:
Touch Screen LCD
Applications:
Cell Culture, Analytical Chemistry, Molecular Biology, Clinical Laboratories
Power Supply:
AC240V-220V / 50-60Hz,110V/60Hz
Product:
Laboratory Water Purification System With EDI
Water Requirement:
Tap Water(TDS<<2000us/cm)
Output Capacity:
30L/hour (Class III),15 L/hour (Class II), 1.5~2.0L/min (Class I)
Dispense Volume:
100-60000ml
Water Tank Capacity:
30L
Control System:
Touch Screen LCD
Applications:
Cell Culture, Analytical Chemistry, Molecular Biology, Clinical Laboratories
Power Supply:
AC240V-220V / 50-60Hz,110V/60Hz
Description
30L Laboratory Ultrapure Water System AC240V Lab Water Filtration System
Type 1, 2,3 Pure RO Ultra Pure Water System NovaPure-E15VF with EDI and Bio-filter

The Laboratory Purification System is engineered for resilience and ease. It provides a dependable pure water source for critical applications, featuring robust monitoring and intuitive maintenance to support continuous lab workflows.

  • Simplest operation: An intuitive four-key interface ensures quick, error-free operation. Advanced processes are automated by the microprocessor. PIN protection maintains consistent operational parameters.
  • Rapid cartridge replacement: Quick-connections enable cartridge swaps without tools. Isolated modules can be changed individually. The streamlined cabinet is designed for front-serviceability, even in tight spaces.
  • Reliable work: Continuous feedwater conductivity monitoring acts as an early warning system. Protecting the purification stages from poor input water extends their life and secures final water quality.
  • Conductivity measurement: Trust in high-precision measurement. Each cell is factory-characterized, and the system performs auto-calibration, providing validated purity data for your records.
  • Temperature compensation: Comply with pharmacopeial standards using accurate temperature compensation. The platinum sensor ensures readings are standardized to 25°C, critical for reliable QC.
  • Flexible dispensing: Adapt the dispenser arm to any container. The removable pistol offers portability. Program large volumes or use hands-free pedal operation for sterile workflows.
  • Clear information: The clear display shows essential data: mode, water quality metrics, and component life, enabling proactive system management.
  • Water leakage protection: A safety sensor linked to the main control board minimizes flood risk through automatic shutdown and alerts.
  • Space saving: A compact footprint allows versatile mounting. The remote dispenser extends system reach, delivering water precisely where needed without moving the main unit.
Configuration
TopPure-SeriesNovaPure-E15VF
Pretreatment module
High pressure pump
Reverse osmosis
Dual wavelength UV-lamp
Ultra-purification cartridge
Ultrafiltration cartridge
EDI Module
Point-of-use filter 
Storage tank 30L
Air filter net for tank
UV-lamp sterilizer for water tank
Remote water dispenser with color display
TOC monitoring module/
Bio-filter
Specifications
ModelNovaPure-E15VF
Feed Water Requirement 
SourceTap water
Conductivity*<2000us/cm
Hardness**<450ppm as CaCO3
Pressure0.05~0.5MPa(7-72psi)
Temperature5~40℃
Purification Water(Class III) 
Ionic Rejection>95%
Bacteria Rejection>99%
Conductivity1~20us/cm
Productivity Rate30L/h
High Quality Purification Water(Class II) 
Resistivity At 25℃10MΩ.cm
TOC<30ppb
Dissolved Organic<0.1ppm
Productivity Rate15L/h
Ultrapurification Water(Class I) 
Resistivity At 25℃18.2MΩ.cm
Conductivity At 25℃0.055us/cm
TOC Level**1~5ppb
Endotoxin(Pyrogens)****<0.001EU/ml
Particulate(≥0.02?m)<1pc/ml
Bacteria*<0.1 cfu/ml
Rnase / Dnase**Free
Manual dispense flow rate1.5~2.0L/min
Automatic dispense volume100~60000ml
Electrical Requirements 
Electrical Voltage110V/220V±10%
Electrical Frequency50HZ/60HZ
Packing Information 
Net Weight 
Main units35kg
Water tank (30L)7kg
External Dimensions(WxDxH) 
Main units315x525x570mm
Water tank (30L)380x380x595mm
Shipping weight 
Main units37kg
Water tank (30L)15kg
Shipping Dimensions(WxDxH) 
Main units525x610x770mm
Water tank (30L)520x440x615mm
Introduction about laboratory water purification system
Advanced Lab Water Purification: Real-Time Monitoring & Smart Safeguards

Modern advanced laboratory water purification systems are engineered for uncompromising quality control through integrated, real-time surveillance. These systems ensure every drop of water meets stringent purity standards required for sensitive applications.
Continuous Quality Verification
At the heart of these systems is proactive monitoring. High-precision resistivity and conductivity meters provide instantaneous, continuous feedback on ionic purity levels. Simultaneously, specialized TOC analyzers meticulously track the concentration of organic impurities, offering a complete picture of water quality.
Connected Intelligence & Data Integrity
Equipped with comprehensive data logging capabilities, these units create a verifiable audit trail of all purity parameters. Support for Lab network connectivity enables remote system oversight from a workstation or office. This connectivity allows for instant email or SMS alerts and notifications, keeping personnel informed of system status in real time.
Proactive Protection for Critical Work
This intelligent surveillance network guarantees that output water consistently adheres to its specified purity grade, such as Type I or Type II. The system is designed with automatic fail-safes: if any parameter falls below a predefined threshold, it can automatically halt water distribution and immediately notify assigned staff. This crucial feature provides proactive protection, effectively preventing the accidental use of non-compliant water in experiments, cell cultures, or analytical procedures, thereby safeguarding research integrity and valuable samples.

Send your inquiry
Please send us your request and we will reply to you as soon as possible.
Send