Product Details
Place of Origin: China
Brand Name: Senova
Certification: CE
Model Number: BioPure30T-RO
Payment & Shipping Terms
Minimum Order Quantity: 1 unit
Price: Available upon request
Packaging Details: plywood case
Delivery Time: 5-20 working days
Payment Terms: T/T
Supply Ability: 100 units
Feed Water Source: |
Tap Water |
Resistivity: |
≥10MΩ.cm (Type II) |
Production Capacity: |
1.0~1.5L/min |
Display: |
LCD Touch Screen |
Storage Tank Capacity: |
30 Liters |
Power Supply: |
220V/50Hz Or 110V/60Hz |
Feed Water Source: |
Tap Water |
Resistivity: |
≥10MΩ.cm (Type II) |
Production Capacity: |
1.0~1.5L/min |
Display: |
LCD Touch Screen |
Storage Tank Capacity: |
30 Liters |
Power Supply: |
220V/50Hz Or 110V/60Hz |
| Model | BioPure30T-RO |
| Productivity Rate | 30L/h |
| Feed Water Requirement | |
| Source | Tap Water |
| Conductivity* | <2000us/cm |
| Hardness** | <450ppm as CaCO3 |
| Pressure | 0.1-0.5MPa |
| Temperature | 5~40℃ |
| High Quality Purification Water (Type II) | |
| Resistivity At 25℃ | ≥10MΩ.cm |
| TOC | <30ppb |
| Dissolved Organic | <0.1ppm |
| Particulate(≥0.02um) | <1pc/ml |
| Bacteria | <1cfu/ml |
| Flow Rate | 1.0~1.5L/min |
| Electrical Requirement | |
| Electrical Voltage | 110V/220V±10% |
| Electrical Frequency | 50Hz/60Hz |
| Packing Information | |
| Net Weight | |
| Main unit | 25Kg |
| Water tank | 5Kg |
| External Dimensions (WxDxH,mm) | |
| Main unit | 315x525x570 |
| Water tank | 380x380x595 |
| Gross Weight | |
| Main unit | 37Kg |
| Water tank | 13Kg |
| Packing size(WxDxH,mm) | |
| Main unit | 525x610x770 |
| Water tank | 520x440x615 |
*When hardness of feed water is high(>450ppm as CaC03), 0.5T water soften tank is recommended
To produce water suitable for scientific work, laboratory purification systems utilize a combination of technologies. Reverse osmosis acts as a preliminary barrier, filtering out large particles and dissolved ions. Deionization further purifies by exchanging ions with resins, thereby increasing water resistivity. UV irradiation effectively inactivates microorganisms and breaks down organic contaminants, reducing total organic carbon (TOC). Additional filtration via activated carbon or ultrafiltration membranes targets specific impurities. This layered approach allows labs to generate on-demand water tailored to specific needs, from basic cleaning to advanced analytical techniques, all while minimizing external contamination risks.